Important News for gas dividers at Be.T.A. Strumentazione home !
Following
some Customer suggestion, we did recently revise the BetaCAP30 series
project .
We did not
touch the most appre-ciated details, than the capillaries bench and the
overall size but we did update the following points :
- The
electronic circuit has been redesigned using a more powerful
microcontroller offering much more program memory, more interface
pins, higher speed, higher A/D converters resolution.
- The
pressure sensors has been changed : no more differential pressure
sensors connected by tubes, but relative pressure sensors inbuilt in the
capillaries manifold with the surface in touch with the gas made by
steel AISI316L. Now the gas divider measure the two inlet
pressures, outlet pressure, barometric pressure and internal temperature
and we also acquire three signals (measuring signals from gas analyzers)
with an isolated interface.
The result
of those modifications (redesigning the firmare) is very important :
- the
pressures (span gas pressure, diluting gas pressure, output diluted gas
pressure, barometric pressure) are now indicated in hPa with 4 digits (±1
mBar) and the resolution of the analog inlets is also improved.
- the ratio between the two inlets
differential pressures is displayed to easily allow the compensation of
input gases different viscosities (just in case of highly concentrated
gases)
- The menus has been redesigned with more
functions and password protected parameters settings for internal
calibrations. The calculation of output diluted concentration now
takes into account the case where the span gas is also included in the
diluting gas,
- The calculation of the output diluted
concentration, now may be corrected to neutralize the small errors
indicated in the metrological certificate |